ShinExcita · Website Preview

KE200-U Rack Gas Analyzer

Select the KE200-U configuration by the measured gas or moisture requirement. Each configuration below retains its own measurement range and sample conditions.

Discuss this equipment

Technical details

KE200-U Online Oxygen Analyzer

A 3U rack instrument for continuous oxygen measurement, with a selectable range.

O₂ range upper limits
10, 100 or 1000 ppm, or 1%, 5%, 25% or 100%
Sample flow
400 ± 50 mL/min
Response time
T₉₀ ≤ 50 s
Instrument size
483 × 137 × 350 mm, W × H × D
KE200-U Hydrogen Analyzer

KE200-U Hydrogen Analyzer

Select the sensing principle and the hydrogen range together. Thermal conductivity and electrochemical versions cover different concentrations.

Thermal conductivity range options
0–30%, 0–50% or 0–99.99% H₂
Electrochemical range options
0–30, 0–100 or 0–2000 ppm H₂
Sample flow
400–800 mL/min
Response time
T₉₀ ≤ 60 s

KE200-U Infrared Gas Analyzer

A rack instrument that uses infrared sensing. Specify the target component and concentration range when selecting the version.

Sample temperature
0 to +50 °C
Sample flow
400–800 mL/min
Response time
T₉₀ ≤ 30 s
Analog output
4–20 mA DC, non-isolated, load <500 Ω

KE200-U Online Dew Point Analyzer

Compare the standard and extended dew point ranges and specify the required analog or serial interface.

Dew point options
−80 to +20 °C standard, −100 to +20 °C extended
Sample flow
2–4 L/min
Inlet pressure
0.05–0.35 MPa gauge
Serial output
RS232 standard or RS485

Selecting a configuration

What to send with your inquiry

Choose the instrument by the measured gas, background gas, range and sample conditions. Compare flow, pressure and signal requirements before selecting a configuration.

Contact ShinExcita